Chapter 1 Safety
WARNING:
X
Electric shock may exist result if this system, including and all of its externally mounted recording
and monitoring devices, is not properly grounded.
X
Do not remove the covers on the system; hazardous voltages are present inside. Cabinet panels
must be in place while the system is in use. All internal adjustments and replacements must be
made by a qualified Samsung Medison Customer Service Department.
X
Check the face, housing, and cable before use. Do not use and disconnect the power source, if
the face is cracked, chipped, or torn, the housing is damaged, or if the cable is abraded.
X
Always disconnect the system from the wall outlet prior to cleaning the system.
X
All patient contact devices, such as probes and ECG leads, must be removed from the patient
prior to application of a high voltage defibrillation pulse.
X
The use of flammable anesthetic gas or oxidizing gases (N
2
O) should be avoided.
X
Avoid places where the system is likely to be difficult to operate the disconnection device.
X
Do not use HF surgical equipment with the system. Any malfunctions in the HF surgical
equipment may result in burns to the patient.
X
The System must only be connected to a supply mains with protective earth to avoid risk of
electric shock.
CAUTION:
X
The system has been designed for 100-120VAC and 200-240VAC; you should select the input
voltage of printer and VCR. Prior to connecting a peripheral power cord, verify that the voltage
indicated on the power cord matches the voltage rating of the peripheral device.
X
An isolation transformer protects the system from power surges. The isolation transformer
continues to operate when the system is in standby.
X
Do not immerse the cable in liquids. Cables are not waterproof.
X
The auxiliary socket outlets installed on this system are rated 100-120V and 200-240V with
maximum total load of 200W. Use these outlets only for supplying power to equipment that is
intended to be part of the ultrasound system. Do not connect additional multiple-socket outlets
or extension cords to the system.
X
Do not connect any peripheral devices not listed in this manual to the auxiliary socket outlets of
the system.
X
Do not touch SIP/SOP and the patient simultaneously. There is a risk of electric shock from
leakage current.