Chapter 5 Measurements And Calculations Measurement Accuracy ...
5 - 10 SONOACE X6 Operation manual ▐ Basic Measurement Operations The following is the information on common button operations for basic measuremen
Chapter 5.Measurements and Calculations 5 - 11 Distance Measurement ▐ 2D Distance (Line) This is a basic measurement that is available in all diagn
5 - 12 SONOACE X6 Operation manual 3. With the Trackball, draw a desired curve and press the Set button to specify the end point. ※ Tip! Editing
Chapter 5.Measurements and Calculations 5 - 13 Please refer to the table below for Hip Joint Type information: Type α β 1a 60 ≤ α< 90 0 < β< 5
5 - 14 SONOACE X6 Operation manual 2. Specify two points and measure the straight distance between them. The method for measuring a line is the sa
Chapter 5.Measurements and Calculations 5 - 15 specify a point in a Spectral Doppler image and trace a curve from that point to calculate the velocit
5 - 16 SONOACE X6 Operation manual Circumference and Area Measurement ▐ 2D Ellipse This is a basic measurement that is available in all diagnosis
Chapter 5.Measurements and Calculations 5 - 17 ▐ 2D Trace This is a basic measurement that is available in all diagnosis modes. You can measure the c
5 - 18 SONOACE X6 Operation manual Volume Measurement ▐ 3 Distance This is a basic measurement that is available in all diagnosis modes. You can me
Chapter 5.Measurements and Calculations 5 - 19 4. A new measurement cursor appears. When more than 4 straight lines are specified, the lines are del
5 - 2 SONOACE X6 Operation manual Fetal Echo Calculations ...
5 - 20 SONOACE X6 Operation manual Calculations by Application Press the Calculator button on the control panel. Things to note ▐ Before Taking
Chapter 5.Measurements and Calculations 5 - 21 In case of measuring distance, choose a measurement method from Line or Trace. When circumference and
5 - 22 SONOACE X6 Operation manual Common Measurement Methods This section provides information on the common measurement methods used for applicat
Chapter 5.Measurements and Calculations 5 - 23 - If the Wall Filter is set too high, only part of the spectrum is displayed. In this case, Trace can
5 - 24 SONOACE X6 Operation manual Measurement items for Doppler Spectrum are as follows: Item Type Unit Equation PSV (Peak Systolic Velocity
Chapter 5.Measurements and Calculations 5 - 25 1. Select the %StA menu and the first cursor will appear in 2D Mode. 2. Measure the area of the vess
5 - 26 SONOACE X6 Operation manual OB Calculations ▐ Before Obtaining OB Calculations OB Basic Information Type the information needed for OB di
Chapter 5.Measurements and Calculations 5 - 27 Measured items are automatically recorded in a report. [Figure 5.1 OB Measurement Menu] NOTE For i
5 - 28 SONOACE X6 Operation manual How to Measure a Distance Measurement Item BPD: Measure the maximum transverse diameter value when symmetri
Chapter 5.Measurements and Calculations 5 - 29 Measurement Item Position (Measurement points) BPD, OFD, HC –fetal head APD, TAD, APTD, TTD – fet
Chapter 5.Measurements and Calculations 5 - 3 Measurement Accuracy Measurement values can vary, depending on the nature of the ultrasound, the body’
5 - 30 SONOACE X6 Operation manual LV – fetal Vertebrae In a report, a maximum of three mean average of new GA values and recently measured val
Chapter 5.Measurements and Calculations 5 - 31 to the measurement mode. 4. The measurement values taken from each plane are added and displayed o
5 - 32 SONOACE X6 Operation manual GYN Calculations ▐ Before Obtaining GYN Calculations Patient information needed for GYN diagnosis can be inputte
Chapter 5.Measurements and Calculations 5 - 33 ▐ GYN Measurements in 2D Mode Menu Sub menu Input Meas. Uterus Uterus Length Distance Uterus Height
5 - 34 SONOACE X6 Operation manual Cardiac Calculations NOTE Since the cardiac blood flow is fast, m/s is used as a velocity unit. ▐ Before Obt
Chapter 5.Measurements and Calculations 5 - 35 [Figure 5.3 Cardiac Measurement Menu] ▐ Cardiac Measurements in 2D Mode Menu Sub menu Input M
5 - 36 SONOACE X6 Operation manual Menu Sub menu Input Meas. Ventricle) Mass Endo. Area (Endocardial) Area LV Length (Left Ventricle Length) Di
Chapter 5.Measurements and Calculations 5 - 37 in the case of the diastole of Left Ventricle. - Myocardial thickness = SQRT(Dia.A / Pi) – SQRT(Sys
5 - 38 SONOACE X6 Operation manual Menu Sub menu Input Meas. All Points (D-E Excursion, E-F slope, A-C Interval, E Point Septal Separation) Same
Chapter 5.Measurements and Calculations 5 - 39 - Set the start and end points of the spectral waveform with a bar. - Rotate the Menu / Angle dial-b
5 - 4 SONOACE X6 Operation manual ▐ Doppler Signal Adjustment - During velocity measurement, an error may occur depending on the cosine angle betw
5 - 40 SONOACE X6 Operation manual ▐ Cardiac Measurements in Spectral Doppler Mode Menu Sub menu Input Meas. MV (Mitral Valve) Inflow Trace –Limi
Chapter 5.Measurements and Calculations 5 - 41 Menu Sub menu Input Meas. Acc (Acceleration Time) Time or Calculated from Trace Dec (Deceleration T
5 - 42 SONOACE X6 Operation manual Menu Sub menu Input Meas. Hepatic Vein All N/A Dias Vel. (Diastolic Velocity) Velocity Sys Vel. (Systolic Ve
Chapter 5.Measurements and Calculations 5 - 43 - A. Rev Dur. measures the time duration by measuring the distance between the Mitral Valve Closing
5 - 44 SONOACE X6 Operation manual Vascular Calculations ▐ Before Obtaining Vascular Calculations Set the related menus for convenient measuremen
Chapter 5.Measurements and Calculations 5 - 45 ▐ Vascular Measurements Menu Sub menu Input Meas. Rt./ Lt. Subclavian Trace – Auto, Limited, Manua
5 - 46 SONOACE X6 Operation manual ▐ Volume Flow Measurements The Volume Flow menu is shown below. For more information, refer to ‘Volume Flow Mea
Chapter 5.Measurements and Calculations 5 - 47 Urology Calculations ▐ Before Obtaining Urology Calculations For more convenient measurement, set the
5 - 48 SONOACE X6 Operation manual The measurement of Transitional Zone Prostate Volume, Bladder Volume, Left Kidney Volume, and Right Kidney Vo
Chapter 5.Measurements and Calculations 5 - 49 If Ellipsoid is set in Setup, Main Diameter and Beside Diameter values can be obtained in order.
Chapter 5.Measurements and Calculations 5 - 5 Optimization of Measurement Accuracy ▐ 2D Mode - Resolution is in proportion to the frequency of the p
5 - 50 SONOACE X6 Operation manual ▐ Volume Flow Measurements The Volume Flow menu is shown below. For more information, refer to ‘Volume Flow Mea
Chapter 5.Measurements and Calculations 5 - 51 Fetal Echo Calculations ▐ Before Obtaining Fetal Echo Calculations Set the related menus for convenien
5 - 52 SONOACE X6 Operation manual ▐ Fetal Echo Measurements in 2D Mode Menu Sub menu Input Meas. 2D Echo Asc. Aorta (Ascending Aorta) Distance
Chapter 5.Measurements and Calculations 5 - 53 ▐ Fetal Echo Measurements in M Mode Menu Sub menu Input Meas. Fetal M Mode IVSd (Inter Ventricular S
5 - 54 SONOACE X6 Operation manual ▐ Fetal Echo Measurements in Spectral Doppler Mode Menu Sub menu Input Meas. MPA (Main Pulmonary Artery) Trace
Chapter 5.Measurements and Calculations 5 - 55 - Peak E measures the Ventricular Systolic peak. - Peak A measures the Atrial Systolic peak.
5 - 56 SONOACE X6 Operation manual Report Check the measurement and calculation results in the report of each application. Press the Report b
Chapter 5.Measurements and Calculations 5 - 57 Editing Report Press the Edit to edit the specific information. After editing, press the Ok at the t
5 - 58 SONOACE X6 Operation manual Set the computer or printer, connects to this system, like following; - 9,600 Baud - 8-bit data - No pa
5 - 6 SONOACE X6 Operation manual - It is not recommended to use images in Color/Power Doppler Mode for measurement of accurate blood flow velocit
Chapter 5.Measurements and Calculations 5 - 7 Measurement Accuracy Table The following tables show the accuracy of the measurements available using t
5 - 8 SONOACE X6 Operation manual (2) Distance is displayed with ‘cm’ and with the second digit. ▐ M Mode Measurements: System Tolerance: (whichev
Chapter 5.Measurements and Calculations 5 - 9 Basic Measurements Press the Caliper button on the control panel. NOTE Take basic measurements of di