Seitenansicht 29
7-4
Operation Manual
Basic Use
The items that are commonly used in each diagnosis mode are shown below:
Using the Control Panel
The items that can be used in each diagnosis mode are provided as menu items. You can change the
image format or optimize an image to facilitate your diagnosis.
Gain
Use the dial-button on the control panel. The Gain button appears differently depending on the
diagnosis mode, but it is usually in the form of a dial-button used to select a diagnosis mode.
You can adjust the brightness of an image. If you rotate the Gain dial-button clockwise, its value
increases.
TGC (Time Gain Compensation)
Tap the TGC button on the touch screen to open the TGC window for use.
In general, ultrasound penetration gets weaker with depth. TGC can be used to compensate for this
effect.
The product provides eight TGC sliders for varying depths, allowing you to adjust Gain by area.
Among the eight sliders, the top slider represents the shallowest area, while the lower sliders
represent the deeper ones.
Move the slider to the right to increase Gain, brightening the image.
When the TGC preset is selected, the TGC preset is applied to the system and then the system is
switched to diagnosis mode. The user can adjust TGC as needed, and press Save to save it for use.
Focus
Use the Focus switch on the control panel to adjust the location of the focus.
Depth
Use the Depth switch on the control panel to adjust the image scanning depth. The allowable
range for adjustment varies depending on the probe type.
Zoom
Use the Zoom dial-button on the control panel.
You can magnify an image. An image can be magnified by either Read Zoom or Write Zoom.