Color/Power Doppler Mode
▶
A protocol is not specied for images in Color Doppler Mode or Power Doppler Mode. Therefore, the
same limitations imposed when measurements are taken in B/W images apply to the accuracy of the
measurements taken in these modes.
▶
It is not recommended to use images in Color/Power Doppler Mode for measurement of accurate
blood ow velocity.
▶
The amount of blood ow is calculated based on the average velocity rather than the peak velocity.
▶
In all applications, the amount of blood ow is measured in PW Spectral Doppler Mode.
Cursor Position
▶
All measurements are aected by input data.
▶
To ensure accurate positioning of the cursor:
Adjust the images on the screen so that they are displayed at maximum granularity.
Use the front edge or boundary point of a probe to make the start and end points of a measurement
object more distinct.
Make sure that the probe direction is always aligned during measurement.