Chapter 7 Utilities
7-29
Conguration Info
Specify the unique value for a printer. Please refer to the DICOM Conformance Statement for the printer.
[Figure 7.20 DICOM Configuration - Print]
Worklist Server Settings
Select WORKLIST under Services. Congure the Modality Worklist Service using DICOM.
Open MWL on Patient key
Sets the screen that appears when pressing the Patient button on the control panel.
If this function is selected, pressing the Patient button opens the Modality Worklist window. If not se-
lected, pressing the Patient button opens the Patient Information screen.
Update Method
Sets the method for updating Worklist.
Only on user Request: Update only when asked by the user.
Tips!
To update a worklist, in the Search tab on the Patient Information screen, select Worklist for
Search Source and press Search.
On Startup and Every: Worklist is updated at the system boot-up and then updated continually at
a set interval. Note that Worklist is not updated while the Modality Worklist window is open.