1-26
User Manual
Probes
Each probe model available has unique specications for the contact area, beam shape, and
center frequency. Settings are reset when you select a probe. Samsung Medisons factory defaults
vary with probe, application and mode. Defaults that are below the FDA limits have been chosen
for intended use.
Depth
An increase in the 2D-mode depth will automatically decrease the 2D-mode frame rate. This
would decrease the TI. The system may also automatically choose a deeper 2D-mode focal depth.
A change of focal depth may change the MI. The MI displayed is that of the zone with the largest
peak intensity.
Application
Acoustic output defaults are set when you select an application. Samsung Medison’s factory
defaults vary with probe, application, and mode. Defaults that are below the FDA limits have been
chosen for intended use.
Related Guidance Documents
For more information about ultrasonic bioeects and related topics, refer to the following:
X
Medical Ultrasound Safety (AIUM, 2009). (A copy of this AIUM Clinical User Education Brochure
is shipped with each system.)
X
AIUM Consensus Report on Potential Bioeects of Diagnostic Ultrasound: Executive Summary,
J. Ultrasound in Medicine, 2008, Vol. 27, Num. 4.
X
WFUMB. Symposium on Safety of Ultrasound in Medicine: Conclusions and Recommendations
on Thermal and Non-thermal Mechanisms for Biological Eects. Ultrasound in Med. & Biol;
1998, 24: Supplement 1.
X
Bioeects and Safety of Diagnostic Ultrasound (AIUM, 1993)
X
Guidelines for the safe use of diagnostic ultrasound equipment. (BMUS, 2009)
X
Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems
and Transducers (U.S. FDA – 2008)
X
Particular requirements for the basic safety and essential performance of ultrasonic medical
diagnostic and monitoring equipment. (IEC, 2007)