Samsung PT60A Bedienungsanleitung Seite 48

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 232
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 47
1-30
User Manual
prr PULSE REPETITION RATE, the inverse number of the time interval between two
contiguous acoustic pulse (unit: Hz)
TIB BONE THERMAL INDEX, a thermal index for a focal zone formed near a bone
after the ultrasound beam passes through soft tissue, e.g. applied to a fetus (2nd
or 3rd trimester) or to the head of a neonate (through the fontanel) (unit: N/A)
TIC CRANIAL BONE THERMAL INDEX, a thermal index for an ultrasound beam
entering the body and passing through a bone, e.g. skull of children or adults
(unit: N/A)
TIS
scan
Soft tissue thermal index in scanning mode (unit: N/A)
TIS
non-scan
Soft tissue thermal index in non-scanning mode (unit: N/A)
t
d
PULSE DURATION (unit: us)
z
_at_max_Ipi,α The location where PULSE-INTENSITY INTEGRAL is maximal (unit: cm)
z
b
DEPTH FOR BONE THERMAL INDEX (unit: cm)
z
bp
BREAK-POINT DEPTH, which is EQUIVALENT APERTURE DIAMETER multiplied by
1.5 (unit: cm)
z
s
DEPTH FOR SOFT-TISSUE THERMAL INDEX, the distance from a plane where the
product of minimum attenuated output power, ATTENUATED SPATIAL-PEAK
TEMPORAL-AVERAGE INTENSITY, and 1cm
2
is maximal at the distance range
that is equal to, or greater than, the equivalent aperture diameter multiplied
by 1.5, when the beam dimension of -12dB output is dened along the beam
alignment axis (unit: cm)
Acoustic Measurement Precision and Uncertainty
The Acoustic Measurement Precision and Acoustic Measurement Uncertainty are described below.
Quantity Precision Total Uncertainty
Ipi,α (attenuated pulse intensity integral)
3.2% +21% to -24%
P (acoustic power)
6.2% +/- 19%
P
r,α
(attenuated rarefaction pressure)
5.4% +/- 15%
f
awf
(acoustic working frequency)
< 1% +/- 4.5%