5-30
Operation Manual
OB Calculations
NOTE:
X
Ductus Venosus and Fetal HR can only be measured in Doppler Mode.
X
For information on basic measurement methods, refer to the ‘Basic Measurement and ‘Common
Measurement Methods’ sections in this chapter.
X
For references on measurement items, see the ‘Reference Manual - Part 1’.
Before Taking OB Measurements
OB Basic Information
Enter the information required for OB diagnosis in the Patient Information window. The basic OB
information includes LMP (Last Menstrual Period) and Gestations.
Once LMP is entered, EDD (Estimated Delivery Date) and GA (Gestational Age) are calculated
automatically. LMP is required for the calculation of values such as EDD and SD in obstetrics
measurement.
X
EDD(LMP) = LMP + 280days
X
GA(LMP) = Current System Date - LMP
Regardless of LMP, enter the EDD with a physicians opinion into Estab. Due Date. If LMP is not
available, when Estab. Due Date is modied, LMP is automatically calculated, and the ‘C’ mark is
displayed next to the LMP information.
A maximum of four fetuses can be entered in the Gestations menu. The default value is ‘1’. In the
case of twins, enter ‘2’.
For further information about patient information menus and how to input patient information,
refer to the Patient Information Entry section in Chapter 3. Starting Diagnosis’.