Measurement Menu Item Mode Method Unit
General
Auto Trace PW Doppler spectrum trace cm, mm
Limited Trace PW Doppler spectrum trace cm, mm
Manual Trace PW Doppler spectrum trace cm, mm
PSV PW Velocity cm/s, m/s
EDV PW Velocity cm/s, m/s
%StA All Calculation after area measurement %
%StD All
Calculation after distance
measurement
%
Vesl.Area All Area cm
2
, mm
2
Vesl.Dist. All Distance cm, mm
The measurement methods for Transitional Zone Prostate Volume, Bladder Volume, Left Renal Volume,
Right Renal Volume are the same as for Prostate Volume.
3 Distance * Factor
The measurement method is identical to ‘3 Distance. The volume is calculated by using the factor
value.
Volume (ml) = Factor x Distance1 x Distance2 X Distance3