
5 - 38 SONOACE X6 Operation manual
Menu Sub menu Input Meas.
All Points (D-E Excursion, E-F slope, A-C Interval,
E Point Septal Separation)
Same as above
Heart Rate
Heart Rate
Time interval & no. of
beats
■ Lt. Ventricle
- The sub-menus of Lt. Ventricle are measured in the same way as in 2D Measure.
- RVDd is measured in the same way as in LVDd.
■ Ao/LA
The sub-menus of Ao/LA are measured as shown below.
- Ao Root Dia measures the distance between the Anterior Aortic Wall and
Posterior Aortic Wall.
- AoV Cusp Sep. measures the distance between the Coronary Cusp and Non
Coronary Cusp.
- LA Dia measures the distance between the Posterior Aortic Wall and Left Arterial
Wall.
- LV PET measures the distance between the ECG’s Q Wave and Aortic Valve
Opening.
- LV ET measures the distance between the Aortic Valve Opening and Aortic Valve
Closing.
■ MV
The sub-menus of MV are measured as shown below.
- A-C Interval calculates the time duration by measuring the distance between the A
point and C point in the Mitral Valve measurement screen.
- D-E measures the distance between the E point of the Anterior Mitral Valve and
the D point of the Mitral Valve Opening.
- E-F Slope calculates the velocity by measuring the distance between the E point
and F point in the Mitral Valve measurement screen.
- EPSS measures the distance between the E point in the Mitral Valve and Posterior
Interventricular Septum. The user can measure only in the upper direction of the E
point, and only in the lower direction in Apex.
■ Heart Rate
Select Heart Rate, and heart rate is calculated.
Kommentare zu diesen Handbüchern