5 - 4 SONOACE X6 Operation manual
Doppler Signal Adjustment
- During velocity measurement, an error may occur depending on the cosine angle
between the blood flow and the ultrasound beam.
- For Doppler velocity measurements, the most accurate results can be ensured when the
ultrasound beam is aligned in parallel with the blood flow.
- If that is not possible, the angle between them should be adjusted by using the Angle
option.
Aliasing
- PW Spectral Doppler Mode uses a signal sampling technique to calculate the frequency
(or velocity) spectrum.
- Adjust the baseline or the velocity scale to minimize aliasing. A lower frequency probe
can also be used to reduce aliasing.
- Aliasing is dramatically reduced in CW Spectral Doppler Mode.
Calculation Equation
- Some of the calculation equations used for clinical purposes originate from hypotheses
and approximation.
- All calculation equations are based on medical reports and articles.
Human Error
- Human error may occur due to inappropriate use or lack of experience.
- This can be minimized through compliance with and thorough understanding of the
manuals.