Chapter 1 Safety
CAUTION:
X
The system has been designed for 100-240VAC; you should select the input voltage of any
connected printer and VCR. Prior to connecting a peripheral power cord, verify that the voltage
indicated on the power cord matches the voltage rating of the peripheral device.
X
An isolation transformer protects the system from power surges. The isolation transformer
continues to operate when the system is in standby.
X
Do not immerse the power cable in liquids. The power cable is not waterproof.
X
The auxiliary socket outlets installed on this system are rated 100-240VAC with a maximum total
load of 100VA. Use these outlets only for supplying power to equipment that is intended to be
part of the ultrasound system. Do not connect additional multiple-socket outlets or extension
cords to the system.
X
Do not connect any peripheral devices that are not listed in this manual to the auxiliary socket
outlet of the system. It may cause an electrical hazard.
X
Do not touch the SIP/SOP terminal on the product while diagnosing the patient. There is a risk
of electric shock from leakage current.
Additional equipment that is connected to medical electrical equipment must comply with the
respective IEC standards (e.g. IEC 60950/EN 60950 for data processing equipment, IEC 60601-1/EN
60601-1 for medical devices). Furthermore, all the components of the product shall comply with the
requirements for medical electrical systems (IEC 60601-1-1/EN 60601-1-1). Anyone who connects
additional equipment to signal input and output ports of medical electrical equipment must verify that
the equipment complies with IEC 60601-1-1/EN 60601-1-1.