Chapter 4 Diagnosis Modes
Select Auto ROI in the menu. Press the button to turn it on or o. If it is turned on, the Volume Box is
automatically placed over an area that will be converted to a 3D image.
When Auto ROI is used, the following should be considered:
Images can be obtained over the fetal body only.
The resulting 3D image is aected by the brightness or contrast of a 2D image.
Once Auto ROI is turned on, the location or size of the Volume Box cannot be changed.
Scan Angle
Use the
Soft Menu
dial-button
[6]
. Rotate the dial-button to adjust the scan angle. The desired scan
angle varies depending on the probe used.
Acquiring 3D Images
1.
Select the desired 3D Mode in the
Soft Menu
[1]
.
2.
Specify the location and size of the ROI box as desired.
3.
Use the
Menu
button and the
Soft Menu
dial-buttons
[2]
-
[6]
to set various parameters.
4.
Once done, press the Freeze or Set button. The system will start acquiring 3D images.
5.
Once 3D images are acquired, the
3D View
or
3D XI
screen appears.
X In the
Soft Menu
[2]
, if
MPR
is selected, 3D View is enabled, and if
MSV
or
Oblique
View
is selected,
3D XI is enabled. Depending on the mode selected, 3D View or 3D XI is displayed in the upper left
corner of the screen.
X If a 3D image is acquired with its left and right sides ipped, it appears in 3D View or 3D XI with its
sides ipped.
6.
After optimizing the acquired 3D images, proceed with diagnosis