8-21
Chapter 8 Measurements and Calculations
Auto Trace
Spectrum is traced automatically. This can be enabled in the measurement menu in Spectral
Doppler mode.
1. Press Auto Trace in the Measurement menu.
2. The system traces the spectrum automatically.
3. When the Trace is complete, the measurement results are displayed on the screen.
Tips!
Things to Consider when using Doppler Spectrum Auto Trace
The state of a Doppler Spectrum may affect the measurement results. Please pay attention to the
following:
X
Causes for Trace Failure
If the Gain is changed while a Doppler image is in Freeze state, Contour Trace and Peak
Trace will not work.
If there is little or no noise in an image without a spectrum, Contour Trace will not work.
If there is severe noise in an image, Contour Trace will not work.
If the Clutter Filter is set too high, Auto Trace and Limited Trace may not work.
X
Possible Causes of Inaccurate Peak Trace
If the PRF (Pulse Repetition Frequency) is lower than the velocity at the site being examined,
aliasing may occur. If the original signals are separated from aliasing, a Trace can be carried
out, but the peak measurement may not be accurate.
If the peak of a spectral waveform is not clear or occurs intermittently, a Trace can be carried
out, but the peak measurement may not be accurate.
If the Doppler Gain is set too high or too low, it becomes difficult to distinguish spectra. This
may result in measurement error.
If the Wall Filter is set too high, the spectrum will only be partially displayed. In this case, a
Trace can be carried out, but Peak measurement may not be accurate.
If abnormal noise or artifacts occurs, a Trace can be carried out, but Peak measurement may
not be accurate.
X
Other
Use of the CW Probe may result in measurement error(s).
Limited Trace is supported only for two-peak spectrums, such as Mitral Valve Inflow and
Tricuspid Valve Inflow in the cardiology application.