Chapter 8 Measurements and Calculations
4. Use the trackball and the Set button to set a location for IMT measurement.
X Measure the length of a specic section. Press and hold the Set button at the starting point
and use the trackball to specify the end point.
X You can change the Near and Far positions of the measured area by using the Change button
on the control panel.
5. Once the measurement nishes, measurement values will be listed in a table.
6. Use the touch screen, or the dial-button to select the measurement label, the location of the
blood vessel, and the direction of blood ow.
Setting Up Measurement Items
On the touch screen menu, you may set up the measurement item, the location of the blood
vessel, and the direction of blood ow.
X
Measurement item: Select measurement items among CCA, Bulb, ICA, and ECA. Use the
touch screen button.
X
Direction: Select the blood ow direction either left or right. Use dial-button 2.
X
Location: Select Prox(Proximal), Mid(Middle), or Dist(Distal) as the location of the blood
vessel to measure. Use dial-button 3.
NOTE: Exit Auto IMT+ using following functions:
X
Change Image Mode and Zoom
X
Annotation, BodyMarker, Patient, SonoView, Report, Setup, Help
Measurement Result
X
Max: The maximum thickness of the Intima/Adventitia pair.
X
Mean: The average thickness of the Intima/Adventitia pair.
X
SD: Standard Deviation
X
QI: The distance ratio of the measured point in a distance for Quality Index measurement.
NOTE: Location of measurement result window cannot be changed and it is displayed in the
position established before Auto IMT+ measurement.