Tag Width: It is available when Color Tag is on. Use the Menu dial-button to select the range of
tags, from between 8 - 256. Tag Width aect to Tag Pos.
NOTE: For information on other menu items, please refer to “2D Mode.”
C Mode Softmenu
Following items are displayed on soft menu in C mode. Select the item and adjust the value with the
Softmenu button [1] – [5] on the control panel.
■
Scale
Use the Softmenu button [1]. Pressing the upside of the button makes the PRF (Pulse Repetition
Frequency) increase so that the speed range of the blood ow is widened, and vice versa.
■
Baseline
Use the Softmenu button [2]. When the display method for Color Mode is set to Var, the baseline cannot
be adjusted.
■
Filter
Use the Softmenu button [3]. Filter is an electrical lter used to eliminate low-frequency Doppler signals
caused by the motion of vessel walls. Adjust Cuto Frequency to remove from the screen the Doppler
signals for which the frequencies are lower than the Cuto Frequency. Press the upside of the button to
raise cuto frequency, and vice versa.
■
Steer
NOTE: This item appears in the softmenu only when a Linear Probe is used.
Use the Softmenu button [5]. Adjust the angle of the ultrasound beam. This can minimize loss of color
information based on the angle of the ultrasound beam. Press the button to select Left, None, or Right
for ROI.
NOTE: For information on other softmenu items, please refer to “2D Mode.”