Samsung SONOACE R3 Bedienungsanleitung Seite 29

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 442
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 28
Chapter 1 Safety
1-11
Approved Cables, Probes and Peripherals for EMC
Cables
Cables connected to this product may aect its emissions. Refer to the table below for recommended
cable types and lengths:
Cable Type Length
DVI Shielded Normal
Parallel Shielded Normal
RS232C Shielded Normal
USB Shielded Normal
LAN(RJ45) Twisted pair Any
S-Video Shielded Normal
Foot Switch Shielded 2.5m
B/W Printer Unshielded Coaxial Normal
Printer Remote Unshielded Any
Audio R.L Shielded Normal
VHS Shielded Normal
Probes
The image probe used with this product may aect its emission. The probe listed in ‘Chapter 9. Probes
when used with this product, have been tested to comply with the group1 Class A emission as required
by International Standard CISPR 11.
Peripherals
Peripherals used with this product may aect its emissions.
CAUTION: When connecting other customer-supplied accessories to the system, it is the users
responsibility to ensure the electromagnetic compatibility of the system.
WARNING: The use of cables, probes, and peripherals other than those specified, may result in
increased emissions or decreased immunity of the Ultrasound System.
Seitenansicht 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 441 442

Kommentare zu diesen Handbüchern

Keine Kommentare