Reference Manual 136
LN5–12/40: Pulsed Doppler &2D+Pulsed Doppler
Index Label MI Scan
TIS TIB
TICNon–scan Non–scan
Non–scan
Aaprt≤1cm² Aaprt>1cm²
Maximum index value
1.480 1.640 1.844 1.036** 2.050 (b)
P
r.3
3.330 – – – – –
W
0
–
2D P1: 3.860
PD P1x1: 59.810
PD P1x1: 70.420 – PD P: 62.500 (b)
min[W
.3
, I
ta.3
]
– – – 43.236 – –
z
1
– – – 1.500 – –
Z
bp
– – – 1.658 – –
Z
sp
0.600 – – – 1.600 –
d
eq
(z
sp
)
– – – – 0.385 –
f
c
5.040
2D : 5.760
PD : 5.370
5.500 5.030 5.060 (b)
X A
aprt
–
2D : 3.050
PD : 1.600
1.600 1.600 1.400 (b)
Y A
aprt
– 0.600 0.600 0.600 0.600 (b)
PD
0.727 – – – – –
PRF
999.000 – – – – –
P
r
_Pii_
max
3.520 – – – – –
d
eq
_Pii_
max
– – – – 0.235 –
FL
x
(cm)
– 5.600 6.500 5.600 – (b)
FL
y
(cm)
– 2.100 2.100 2.100 – (b)
I
pa
.MI_
max
498.000 – – – – –
Control 1
MI
Control 2
TIS_as
Control 3
TIS_as_U
Control 4
TIS_bs
Control 5
TIB_bs
Control 6
TIC_as
Control 1: PulsedDoppler Mode, PD: 5.1MHz, Focus: 1.3Cm, Data No: 5971447–376663
Control 2: 2D+PulsedDoppler Mode, 2D: 7.7MHzPD: 5.6MHz, Focus: 5.6Cm, FR: 20.830, Data No: 6138143–577516
Control 3: PulsedDoppler Mode, PD: 5.6MHz, Focus: 6.5Cm, Data No: 5984835–68016
Control 4: PulsedDoppler Mode, PD: 5.1MHz, Focus: 5.6Cm, Data No: 5975895–744993
Control 5: PulsedDoppler Mode, PD: 5.1MHz, Focus: 4.4Cm, Data No: 5975546–73237
Control 6: PulsedDoppler Mode, PD: 5.1MHz, Focus: 6.5Cm, Data No: 5976233–770942
Patient–Applied Part Temperature Table
Transducer Surface Temperature In–Air (°C) Surface Temperature Simulated–Use (°C)
C2–4/20 <41 <41
C2–5 42.2 <41
C2–8 <41 <41
CN2–8 43.2 <41
CN4–9 <41 <41
EC4–9 <41 41.9
EVN4–9 <41 41.9
L5–12/60 <41 42.4
LE5–12 45.1 41.2
LN5–12 41.1 <41
LN5–12/40 <41 42.3