GE HEALTHCARE
DIRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
4-26 Section 4-3 - General Procedure
4-3-15-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
1.) Insert a formatted CD-R into the drive.
2.) On the Keyboard, press UTILITY
.
3.) On the LCD display, press SYSTEM
.
4.) On the LCD display, select BACKUP/RESTORE
.
NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login
window is displayed. Log on with administrator privileges.
5.) In the Backup list, select Patient Archive, Report* Archive and User Defined Configuration.
6.) In the Media field, select CD-RW.
7.) Select BACKUP
.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.
NOTICE
* This function is not available on LOGIQ Book XP PRO.
Figure 4-19 Backup/Restore Menu
Check here to
backup presets
and configurations
Seitenansicht 131