GE HEALTHCARE
DIRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
3-42 Section 3-9 - Loading the System Software
3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above) (cont’d)
9.) After finish updating system, remove the USB memory stick and press any key to shut down
the system. Refer to Figure 3-41 on page 3-42 .
NOTE: If you don not remove the USB memory stick, the software system loading process repeats
when the system boots up.