GE HEALTHCARE
DIRECTION 5194296-100, REVISION A LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
10-10 Section 10-5 - System Maintenance
10-5-8 Battery Performance Maintenance
It is recommended to do battery performance maintenance one time per year.
Please follow the flow chart below to carry out battery performance maintenance.
NOTE: Disconnect all probes when discharge battery.
NOTE: Discharge the battery to let the system automatically shut down.
Figure 10-1 Flow chart of Battery Performance Maintenance
Battery discharge
Battery charge > 3 hour
Yes
No
Battery power > 90%?
Battery charge > 3 hour
Battery power > 90%?
No
Battery is Ok
Yes
Yes
Discharge Replace battery
No
Yes
Battery is Ok
Replace battery
>45 minutes?
Replace battery
>45 minutes?
No