GE HEALTHCARE
DIRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
2-8 Section 2-3 - Facility Needs
2-3-3-1 Recommended and Alternate Ultrasound Room Layout
Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT
LINEN SUPPLY
EXAM INAT IO N
TAB LE
FILM VIEW ER
CO UN TER TO P
FILE
C AB INET
SE CR E TA R YS O R
DOCTO R’S D ES K
FILM
PR O C ESSIN G
R OOM ,
OVERHEAD
LIG H T S D IMM ER
DEDICATED
POW ER
RECEPTACLE
EXTERNAL
PERIPHERALS
SINK
CO UN TER
TO P
FILM VIEW ER
FILM SUPPLIES
SINK
LINEN SUPPLY
PR OB ES /SUPPLIES
24 IN.
(61 CM)
EXAM INA R IO N
TAB LE
76 IN.
(193 CM)
DOO R
30 IN.
(76 CM)
An 8 by 10 foot M inimal Floor Plan
DEDICATED POW ER
OUTLETS
A 14 by 17 foot R ecommended Floor Plan
G E C AB INET
FOT SO FTW A RE
AND MA NU A LS
S cale : Each square equals one square foot
DOO R
42 IN.
(107 CM)
SU C T IO N LINE
EM ERG ECY
OX Y G EN
24 IN.
(61 CM)
76 IN.
(193 CM)
PAT IENT
TO ILET
FA C ILITY
FOOT
SW
FOOT
SW
LOGIQ Book XP
C O NS O LE
C O NS O LE
18 IN.
(46 CM)
S TOO L
S TOO L
Kommentare zu diesen Handbüchern