GE HEALTHCARE
DIRECTION 5194296-100, REVISION A LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 10 Care & Maintenance 10-3
Section 10-3 Maintenance Task Schedule (cont’d)
NOTE: May require specialized equipment to complete.
NOTE: PMs are not mandatory, the table above is for reference only.
Table 10-2 Customer Care Schedule
Service at Indicated Time Daily Weekly Monthly
Per Facilities
QA Program Notes
Clean Probes •* * or before each use
Clean ECG
•* * or before each use
Inspect AC Mains Cable
• Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean LCD •
Console Leakage Current Checks
See Note
Annually
Peripheral Leakage Current Checks
See Note
Annually
Surface Probe Leakage Current Checks
See Note
Annually
Endocavity Probe Leakage
Current Checks
See Note
Quarterly
Annually
Measurement Accuracy Checks
See Note
Annually
Probe/Phantom Checks
See Note
Quarterly
Annually