Seitenansicht 132
GE HEALTHCARE
D
IRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4-27
4-3-15-3 Restore System Presets and Configurations
1.) Insert the Backup/Restore CD-R into the drive.
2.) On the Keyboard, press UTILITY
.
3.) On the LCD display, press SYSTEM
.
4.) On the LCD display, select BACKUP/RESTORE
.
NOTE: If you are not logged in with administrator privileges, the Operator Login window is displayed.
Log on with administrator privileges.
5.) In the Restore list, select Patient Archive, Report* Archive and User Defined Configuration.
6.) In the Media field, select the Backup/Restore CD-RW.
7.) Select RESTORE
.
The system performs the restore. As it proceeds, status information is displayed on the Backup/Restore
screen.
CAUTION
The restore procedure overwrites the existing database on the local hard drive. Make sure to insert
the correct CD.
NOTICE
* This function is not available on LOGIQ Book XP PRO.
Figure 4-20 Backup/Restore Menu
Check here to
restore
presets and
Seitenansicht 132