Chapter 4 Diagnosis Modes
4-21
Density
Set the scan line density. Use the dial-button to select Medium, High or Low.
Selecting High will increase the number of scan lines and therefore increase the image resolution.
Note that the frame rate will drop.
Baseline
Move the baseline on the color bar. Turn the dial-button clockwise to raise the baseline on the color
bar.
Filter
Press the dial-button to turn it on or o. When on, low-frequency Doppler signals from the
movement of the vessel walls are removed.
Power
Power is the intensity of the acoustic output. Select a value between 10 and 100 with the dial-
button.
Multi Frequency
Adjust the frequency within the ROI box. Select Pen or Gen with the dial-button.
TDI
NOTE:
X
This appears on the soft menu only when using a cardiac application with the phased array
probe.
X
This is available in Color Doppler Mode and Power Doppler Mode.
Press the dial-button to start TDI Mode. TDI stands for Tissue Doppler Imaging. This displays the
movement of tissues such as the heart in color.
To exit TDI Mode and return to Color Doppler Mode, use the dial-button in the soft menu to select
C Mode.