Chapter 5 Measurements and Calculations
5-17
Trace
This is a basic measurement available in all diagnosis modes. You can measure the circumference and
area of an irregular object in a 2D image.
1. Press the Caliper button and select Trace in soft menu [2]. Trace is displayed in the user
information area.
2. Use the trackball and the Set button to specify the start point for tracing over the contour of the
measurement area.
X Use the trackball to place the cursor at the desired position and press the Set button.
3. Trace the curve so that the measurement cursor returns to the start point, and then press the Set
button.
NOTE: Trace lines must be closed. If you press the Set button before tracing is complete, tracing
may be done over a straight line between the current point and the start point, resulting in a
significant error.
4. Once measurement is taken, the result is displayed on the screen.
The following formulae are used for Trace measurements:
, (N = 1, 2… last point)
, (N = 1, 2… last point)
%StA
StA stands for Stenosis Area, which is a basic measurement available in all diagnosis modes. In a 2D
image, the area of a vessel is measured and the stenosis ratio (%) calculated.
1. Press the Caliper button and select %StA in soft menu [2]. “%StA is displayed in the user
information area.
2. Measure the area of the vessel outer wall using the Area measurement method.
3. When a new cursor appears, measure the inner wall area of the vessel under stenosis.
4. %StA is calculated using the following formula:
%StA = (Outer Area – Inner Area) / Outer Area × 100