Chapter 8 Maintenance
8-5
Disinfection
NOTE: Use only recommended disinfectants on system surfaces.
For product disinfection, disinfectants qualied according to the U.S. FDA 510(k) process are
recommended. The following disinfectants are compatible (as qualied according to the FDA 510(k)
process) with Samsung Medison ultrasound products and are thus recommended:
Disinfectant Country Type Active Ingredient FDA 510(k)
Cidex USA Liquid Gluteraldehyde K934434
Cidex Plus USA Liquid Gluteraldehyde K923744
1. Turn o and unplug the product.
2. Prepare a disinfectant based on the mix ratio specied in its user manual.
3. Apply it to product surfaces by following the instructions in its user manual.
4. Air dry or wipe with a dry cloth according to the instructions in its user manual.
Accuracy Check
NOTE: The user must ensure that safety inspections are performed every 2 years according to the
requirements of safety standard EN 60601-1. Only trained persons are allowed to perform these
safety inspections.
The measurements obtained using this product are aected by the product’s condition. Please make sure
to maintain the product in optimal condition in order to obtain reliable measurements.
To keep the product in optimal condition, an accuracy check should be performed annually. The equations
and tables related to measurement accuracy are listed in Chapter 5 “Measurements and Calculations” in
this manual.