Samsung MySono U5 Bedienungsanleitung Seite 135

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 424
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 134
Chapter 4 Diagnosis Modes
4-35
Auto ROI
NOTE: This option appears on the menu only when the application for a 3D probe is set to OB or
OB Early.
Press the dial-button to turn it on or o. When turned on, the volume box is automatically located
in the area that will be converted to a 3D image.
The following should be taken into account with Auto ROI:
X Only images of the fetal body can be acquired.
X It is aected by the brightness or contrast of 2D images
X The volume box cannot be moved or resized.
Load Preset
Select Preset with the dial-button.
When a standard probe is used, available options are Default or from User1 to User5.
Acquiring a 3D/4D Image
1. Specify the location and size of the ROI Box as desired.
2. Adjust the various parameters in the soft menu.
3. Press the Freeze or Set button. The system will start acquiring 3D images.
4. Once a 3D image is acquired, a 3D View screen will be displayed according to the settings. The
current view mode is indicated in yellow on the menu tab on the screen.
X If a horizontally reversed 3D image is obtained, the image will be also shown horizontally
reversed in 3D View or 3D XI.
Seitenansicht 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 423 424

Kommentare zu diesen Handbüchern

Keine Kommentare