Chapter 7 Utilities
7-31
Storage Server Information
Set Services to STORAGE. You can specify an Image Storage Service using DICOM.
■
VOI LUT Setup
Specify VOI LUT (Value Of Interest Look Up Table). Adjust and save the brightness and contrast
of DICOM images. The saved images can be viewed with any PACS device that supports standard
DICOM VOI LUT.
X Window Center: Enter a value for the DICOM Tag (0028,1050). This value represents the
brightness of images displayed by the Storage Service. If it is increased to 128 or higher, the
displayed images become darker. However, this option is available only when it is supported
by the Storage Service.
X Window Width: Enter a value for the DICOM Tag (0028,1051). This value represents the contrast
of images displayed by the Storage Service. If it is increased to 256 or higher, the contrast of the
displayed images decreases. This option is available only when it is supported by the Storage
Service.
■
Include 3D Volume
If this checkbox is checked, 3D images are saved along with 3D volume data. However, these options
can be selected only when the Storage Service supports 3D volume data of Samsung Medison.
[Figure 7.17 DICOM Configuration - Storage]