1-12
Operation Manual
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz
to 80MHz
0.01V
Portable and mobile RF communications
equipment should be used no closer to any part of
the Ultrasound System, including cables, than the
recommended separation distance as calculated
using the equation applicable to the frequency of
the transmitter.
Recommended separation distance
80MHz to 800MHZ
800MHz to 2.5GHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5GHz
3V/m
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recomm-
ended separation distance in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol :
NOTE 1) At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is aected by
absorption and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcasts cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an
electromagnetic site survey should be considered. If the measured eld strength in the location in which the
Ultrasound System is used exceeds the applicable RF compliance level above, the Ultrasound System should
be monitored to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Ultrasound System or using a shielded location with a higher
RF shielding eectiveness and lter attenuation.
b
Over the frequency range 150kHz to 80MHz, eld strengths should be less than [V
1
] V/m.