Chapter 1 Safety
Electrical Safety
This equipment has been veried as a Class I device with Type BF applied parts.
CAUTION:
X
As for US requirement, the LEAKAGE CURRENT might be measured from a center-tapped circuit
when the equipment connects in the United States to 240V supply system.
X
To help assure grounding reliability, connect to a “hospital grade” or “hospital only grounded
power outlet.
Prevention of Electric Shock
WARNING:
X
There is a risk of electric shock if the externally mounted recording and monitoring devices are
not properly grounded.
X
Never open the product housing cover. There is high voltage power within the product. All
internal adjustments and replacements must be made by the Samsung Medison Customer
Service.
X
Always check the product housing, cables, cords, plugs, etc. before use. Do not use and
disconnect the power source, if the face is cracked, chipped, or torn, the housing is damaged, or
if the cable is abraded.
X
Always disconnect the system from the wall outlet prior to cleaning the system.
X
All patient contact devices, such as probes, must be removed from the patient prior to
application of a high voltage defibrillation pulse.
X
Never use the product in proximity to any flammable anesthetic gases(N
2
O) or oxidizing gases.
There is a risk of explosion.
X
Avoid places where the system is likely to be difficult to operate the disconnection device.
X
Be sure only to use an AC adaptor that is recommended by Samsung Medison.
X
Do not use HF surgical equipment with the system. Any malfunctions in the HF surgical
equipment may result in burns to the patient.
CAUTION:
X
An isolation transformer protects the system from power surges. The isolation transformer
continues to operate when the system is on standby.
X
Do not immerse the power cables in liquid. Cables are not waterproof.
X
Do not touch SIP/SOP and the patient simultaneously. There is a risk of electric shock from
leakage current.