Operation Manual
5-14
D A/B
This is a basic measurement available in Spectral Doppler Mode only. You can specify two points in a
Spectral Doppler image and measure the velocity at each point to calculate the ratio of the velocity
between them.
1. Press the Caliper button and select D A/B in soft menu [1]. “D A/B” is displayed in the user
information area.
2. Specify two points for which to measure velocity.
X Use the trackball to place the cursor at the desired position and press the Set button.
3. Once measurement is taken, the result is displayed on the screen.
X V1: Velocity at Point 1
X V2: Velocity at Point 2
X PGmax: Max Pressure Gradient
X V1/V2: Ratio of Velocity
NOTE: If Application is set to Cardiac at Utility > Measure Setup > General > Caliper, the results
including Vmax, PGmax, PGmean, VTI, PHT, Acc, AccT, Dec and DecT will be shown on the screen.
D Trace
This is a basic measurement available in Spectral Doppler Mode only. You can specify a point in a
Spectral Doppler image and trace a curve from that point to calculate the velocity, integral value and
average velocity of blood ow.
1. Press the Caliper button and select D Trace soft menu [1]. “D Trace is displayed in the user
information area.
2. Trace a curve. The measuring method is the same as in “Line Trace.
3. Once measurement is taken, the result is displayed on the screen.
X PSV: Peak Systolic Velocity
X EDV: End Diastolic Velocity
X RI: Resistivity Index
X PI: Pulsatility Index
X S/D: Ratio of PSV to EDV
X Vmean: Average Velocity