1-30
Operation Manual
The document includes analysis and discussion of the sources of error/measurement eects due to:
Balance system calibration.
Absorbing (or reecting) target suspension mechanisms.
Linearity of the balance system.
Extrapolation to the moment of switching the ultrasonic transducer (compensation for ringing
and thermal drift).
Target imperfections.
Absorbing (reecting ) target geometry and nite target size.
Target misalignment.
Ultrasonic transducer misalignment.
Water temperature.
Ultrasonic attenuation and acoustic streaming.
Coupling or shielding foil properties.
Plane-wave assumption.
Environmental inuences.
Excitation voltage measurement.
Ultrasonic transducer temperature.
Eects due to nonlinear propagation and saturation loss.
The overall ndings of the analysis give a rough Acoustic Power accuracy gure of +/- 10% for the
frequency range of 1 - 10 MHz.
Training
The operator of this equipment must become thoroughly familiar with the correct use of the equipment
in order to optimize its performance and recognize possible malfunctions. It is recommended that all
users receive proper training before operation. Applications training is available through the Samsung
Medison Customer Service Department, or from one of their worldwide customer service representatives.