Operation Manual
9-22
4. Begin scanning the patient. Adjust the position of patient so that the location for examination
ts into the needle guideline on the screen.
5. Insert the needle into the needle guide. Direct it into the examination location.
6. To keep the needle securely in the needle guide, press down on the opening of the guide with
the index nger during the examination process.
7. When the examination location is reached, take the needle out of the needle guide.
8. Detach the needle guide, adapter and sheath from the probe.
9. Dispose of the components that are not designed for re-use.
Needle Guide Alignment
The alignment of the needle guide displayed on the system is intended to help you to verify whether
the needle and the needle guide are properly installed. This verication must be carried out prior
to the biopsy examination. If the needle is not following the intended path during the verication
process, stop using the biopsy kit and contact Samsung Medison Customer Service for assistance.
Reverberation or tissue artifacts may produce false needle images, which can cause a lack of clarity.
Ensure the needle path is properly aligned with the guideline, and that a correct needle image is used
to locate the needle.
WARNING:
X
The needle used for this alignment verification must not be used for the actual procedure.
Always use a new, sterile needle for each biopsy procedure.
X
To ensure accurate projection of the needle, use a straight, new needle for each alignment
verification procedure.
1. Attach the biopsy kit to the probe.
2. Set the system depth for the procedure to be performed and select the biopsy menu.
3. Immerse the probe into the water bath, and insert the needle into the needle guide.
4. Make sure that the needle image is on the needle guideline. If so, the needle guide is properly
aligned.
5. If the needle image is not on the needle guideline, check the needle guide or the probe adapter.