Chapter 1 Safety
Biological Safety
For more safety information on probes and biopsy, please refer to Chapter 9 ‘Probes.
WARNING:
X
Ultrasound may have adverse effects within cells and may pose a potential threat to the patient.
If there is no medical benefit, minimize the exposure time and maintain the ultrasound wave
output level at low. Please refer to the ALARA principle.
X
Do not use the system if an error message appears on the video display indicating that a
hazardous condition exists. Note the error code, turn off the power to the system, and call your
local Samsung Medison Customer Service.
X
Do not use a system that exhibits erratic or inconsistent updating. Discontinuities in the
scanning sequence are indicative of a hardware failure that should be corrected before use.
X
The system limits the maximum contact temperature to 43 degrees Celsius, and the ultrasonic
wave output conforms to U.S. FDA regulations.
ALARA Principle
Diagnoses performed with ultrasound equipment may utilize the As Low As Reasonably Achievable
principle. The decision as to what is reasonable is left to the judgment and experience of qualied
personnel. No set of rules can be formulated that would be suciently complete to dictate the correct
response for every circumstance. By keeping ultrasound exposure as low as possible, while obtaining
diagnostic images, users can minimize ultrasonic bioeects.
Since the threshold for diagnostic ultrasound bioeects is undetermined, it is the sonographer’s
responsibility to control the total energy transmitted into the patient. The sonographer must reconcile
exposure time with diagnostic image quality. To ensure diagnostic image quality and limit exposure
time, the ultrasound system provides controls that can be manipulated during the exam to optimize the
results of the exam.
The ability of the user to abide by the ALARA principle is important. Advances in diagnostic ultrasound, not
only in the technology but also in the applications of the technology, have resulted in the need for more and
better information to guide the user. The output indices are designed to provide that important information.
There are a number of variables that aect the way in which the output display indices can be used to implement
the ALARA principle. These variables include mass, body size, location of the bone relative to the focal point,
attenuation in the body, and ultrasound exposure time. Exposure time is an especially useful variable, because
the user can control it. The ability to limit the index values over time support the ALARA principle.