Chapter 9 Probes
9-19
Disinfection
Only disinfect vaginal and rectal probes. A 10-6 reduction in pathogens should be reached following
the disinfection procedure in this manual and using Samsung Medison recommended solutions.
WARNING:
X
If a pre-mixed solution is used, be sure to observe the solution expiration date.
X
The type of tissue a probe will contact during use dictates the level of disinfection required for
the device. Ensure that the strength and duration of solution application are appropriate for
disinfection.
CAUTION:
X
Using a non-recommended disinfectant or not following the recommended disinfection
method can damage and/or discolor the probe and will void the probe warranty.
X
Do not immerse probes for longer than one hour, unless they are sterilizable.
X
Only sterilize probes using liquid solutions. Avoid using an autoclave, gas (Ethylene Oxide), or
other non-Samsung Medison-approved methods.
1. Please refer to the user manual of the disinfectant for storage, use, and disposal of the disinfectant.
2. Mix the disinfectant compatible with the probe according to the instructions for solution strength
provided in the user manual.
3. Immerse the probe in the disinfectant as shown in the illustration below.
4. Following the instructions on the disinfectant, rinse the probe after the immersion process is
complete.
5. Air dry the probe or wipe it with a dry cloth.
[Figure 9.1 Disinfecting Probe]